DHI Implanter Pen Sizes - Diameters, Colours and Order Codes
DHI implanter pen sizes are stated as the diameter of the needle. This guide explains the 5 sizes of the Garry DHI Implanter Pen, Ø 0.7 to 1.1 mm, the colour each size carries on the pen cap and the needle hub, which size suits which graft and how to read the order codes.

DHI implanter pen sizes are stated as one number, the diameter of the needle the pen carries. The Garry DHI Implanter Pen is made in 5 sizes, Ø 0.7, 0.8, 0.9, 1.0 and 1.1 mm, and every size has one colour, yellow, blue, red, white and black in that order. The colour sits on the cap of the pen and on the hub of the needle of the same diameter, so a pen and its needle can be matched at a glance. Each size also has its own order code, DHI-PEN-070 to DHI-PEN-110 for the pens and DHI-TIP-070 to DHI-TIP-110 for the needles. This guide explains what the size describes, what each colour means, which size suits which graft, how a team can use the colours at the loading station, how the needle depth is set and held, and how to read and order the codes, with the Garry DHI Implanter Pen as the worked example.
What a DHI implanter pen size describes
A DHI implanter pen is a spring-loaded instrument for direct hair implantation. A graft is loaded into the hollow needle of the pen, the needle opens the recipient site, and the plunger releases the graft as the needle withdraws, so the site is opened and the graft is placed in one motion. Our guide to DHI hair transplantation follows the 3 steps of that motion.
The size of a pen is the diameter of its needle. The Garry needle is published by its bore, the internal channel a graft is loaded into and travels through, and the pen of a given size takes the needle of the same diameter. Pen and needle are 2 separate products with their own order codes, and they share a size and a colour.
2 things are not part of the size. The first is the depth of placement, which is set on the pen and has its own section below. The second is a set of figures that are not published on the site, such as gauge equivalents, wall thickness and outer diameter tolerances. A clinic that needs one of them for a code can ask, and the answer comes in writing.

The 5 sizes, their colours and their order codes
The table lists every size of the Garry DHI Implanter Pen with its colour, the 2 order codes and the grafts each bore is typically chosen for.
| Size | Colour | Pen order code | Needle order code | Typically chosen for |
|---|---|---|---|---|
| Ø 0.7 mm | Yellow | DHI-PEN-070 | DHI-TIP-070 | Single-hair grafts and fine calibers, the finest bore |
| Ø 0.8 mm | Blue | DHI-PEN-080 | DHI-TIP-080 | Single-hair grafts and fine calibers, a fine bore |
| Ø 0.9 mm | Red | DHI-PEN-090 | DHI-TIP-090 | Mixed follicular units, the standard bore |
| Ø 1.0 mm | White | DHI-PEN-100 | DHI-TIP-100 | Fuller multi-hair grafts, a larger bore |
| Ø 1.1 mm | Black | DHI-PEN-110 | DHI-TIP-110 | Coarse calibers and large units, the largest bore |
The colour is chosen with the size and not separately. Each size comes in its one colour only, so a Ø 0.8 mm pen is blue and is not offered in another colour, and the Ø 0.8 mm needle carries the same blue on its hub. A clinic that orders by size therefore receives the colours that belong to those sizes.
The range runs from Ø 0.7 to 1.1 mm. Across the market implanter needles are commonly offered in roughly 0.6 to 1.2 mm, and sizes outside the Garry range or between its published steps are not offered.

Matching the pen size to the graft
The size follows the graft caliber. The bore has to hold the graft securely enough to place it and loosely enough not to compress it. A bore that is too small jams grafts at loading and squeezes them, and a bore that is too large lets the graft sit loose and opens a larger channel than the graft needs. Our guide to implanter needle sizes explains this sizing logic in full.
In the Garry range the 2 fine bores, Ø 0.7 mm and Ø 0.8 mm, are matched to single-hair grafts and fine calibers, typically at the hairline. Ø 0.9 mm is the standard bore for mixed follicular units. Ø 1.0 mm and Ø 1.1 mm are the larger bores for fuller multi-hair grafts and coarse calibers. The final choice follows the measured graft caliber and the plan of the surgeon.
The caliber of a graft is set at extraction, by the punch diameter and the donor hair, and placement has to accept what harvesting produces. So the pen sizes are planned together with the punch diameters. The table sets the 2 instruments side by side, as our guide to what a hair graft is does for the whole instrument table.
| Graft type | Fue Punch inner diameter | DHI pen and needle size |
|---|---|---|
| Finest single-hair units | Ø 0.60 to 0.65 mm | Ø 0.7 mm, yellow |
| Fine single-hair grafts | Ø 0.70 mm | Ø 0.8 mm, blue |
| Mixed follicular units | Ø 0.75 to 0.80 mm | Ø 0.9 mm, red |
| Fuller multi-hair units | Ø 0.85 mm | Ø 1.0 mm, white |
| Multi-hair units and coarse calibers | Ø 0.90 mm | Ø 1.1 mm, black |
The table is a starting point and not a rule. A real case places a range of graft calibers, so a clinic typically runs 2 or 3 pen sizes side by side and sets up each chair with the pen for the caliber being placed. One size covers a whole case only when the graft mix is unusually uniform.
Using the colours at the loading station
Loading is the step that sets DHI apart from placement with forceps. Each graft is loaded into the needle before placement, usually by a technician while the surgeon places with another pen, and several loaded pens stay in rotation so that placement does not wait for loading.
When 2 or 3 sizes are in use at once, the colour tells the size without anyone reading a code. A team can build simple habits on that.
Match before loading. The hub of the needle and the cap of the pen carry the same colour when they are the same size. A red needle belongs in a red pen, and a pair that does not match is visible before a graft is loaded.
One colour for one graft group. Grafts are sorted by caliber after extraction. Each group can be assigned the pen size that suits it, so the fine single-hair grafts for the hairline go to the yellow or blue pens and the fuller units to the white or black ones.
Name the pen by its colour. A surgeon who asks for the next blue pen and a technician who hands it over are speaking about the same size, the Ø 0.8 mm bore.
The colour is an aid at the chair, and the order code on the product label remains the reference. Checking each code against the order when a delivery arrives keeps the 2 in step.
Needle depth, set on the pen and held by the seat
Size and depth are 2 separate decisions. The size is a property of the pen and the needle and is part of their order codes. The depth belongs to the plan of the surgeon and is set on the pen.
On the Garry DHI Implanter Pen the needle depth is adjustable. Placement depth is set by the exposed needle length, the part of the needle that protrudes from the seat of the pen. The positive needle seat holds that setting through needle changes. Each Garry needle is ground and polished for this seating system, so a new needle of the same order code returns the pen to its set exposed length without re-calibration.
Changing to a larger size changes the bore and leaves the depth decision where it was, with the surgeon. Before each session, 3 checks are worth the few seconds they take. The needle should be seated firmly in the pen, the exposed length should match the planned depth and the spring should return cleanly.

Reading the order codes and ordering replacement needles
Every size has 2 order codes, one for the pen and one for the needle. The letters name the product and the digits give the needle diameter in hundredths of a millimetre.
- DHI-PEN-070 is the Ø 0.7 mm pen, yellow, and DHI-TIP-070 is its needle.
- DHI-PEN-090 is the Ø 0.9 mm pen, red, and DHI-TIP-090 is its needle.
- DHI-PEN-110 is the Ø 1.1 mm pen, black, and DHI-TIP-110 is its needle.
The pen and the needle of one size carry the same digits, which makes a pair easy to check on a quotation. Each size is its own catalog entry with its own product page, for example the Garry DHI Implanter Pen 0.9 mm and the Garry DHI Implanter Needle 0.9 mm, and the needles have their own page under DHI Implanter Needle.
Needles are the consumable in DHI work. The Garry needle is reusable and is designed for up to 15,000 penetrations before the tip dulls, and it is replaced at once when the tip shows any damage or when it no longer seats firmly in the pen. Replacement needles are ordered separately, by the DHI-TIP code in the same diameter and colour as the pen. Ordering by code means a reorder names the needle of the first delivery, because each code refers to one defined specification.
A quote request can list any mix of the 5 sizes, each with its own quantity, and what each quoted line includes is stated on the quotation. Pricing is provided by quotation and is not published on the public site, and the minimum order per code, the lead time and the shipping terms are confirmed on the quotation. The fit of pens or needles of another brand is confirmed in writing and is not assumed, because an implanter needle seats in a specific pen system.
Care between sessions
A size only works as long as the pen and the needle are in good condition. Both are supplied non-sterile and are cleaned and steam sterilized at 134 °C before the first use and before every later use. The pen disassembles without tools, and it goes into the sterilizer disassembled so that steam reaches every surface. Clean the parts with a neutral-pH detergent, flush the needle channel before debris dries, keep saline and chlorine-based agents off the steel and check the spring action after reassembly. Our guide to gravity and prevacuum cycles explains why a prevacuum cycle suits a load of this kind.
Questions clinics ask about DHI implanter pen sizes
Which DHI implanter pen size is used for the hairline? The 2 fine bores, Ø 0.7 mm in yellow and Ø 0.8 mm in blue, are matched to single-hair grafts and fine calibers, which is the work typically done at the hairline. The choice follows the measured graft caliber and the plan of the surgeon.
Can I order a size in another colour? No. Each size comes in one colour only, on the pen cap and on the needle hub, because the colour identifies the diameter. The Ø 0.9 mm pen is red and the Ø 1.0 mm pen is white.
How many pen sizes does a clinic need? A clinic typically runs 2 or 3 sizes side by side, because a real case places a range of graft calibers. Send your graft mix and team size and we suggest a set of order codes.
Does the depth have to be set again after a needle change? No. The exposed needle length is adjustable on the pen and the positive needle seat holds the setting, so a new needle of the same order code returns the pen to its set length. Check the seating and the exposed length before each session.
Is there a 0.6 mm or a 1.2 mm pen? No. The Garry DHI Implanter Pen is published in Ø 0.7 to 1.1 mm only. The DHI Implanter Pen questions and answers cover the line in more depth.
Ordering Garry DHI Implanter Pens by size
Send the order codes and quantities you need, or simply your team size and graft mix, through the contact page, and we answer in writing within 24 hours on business days, with a suggested set of sizes where you ask for one. A small first order across the sizes you are considering can be quoted, so your team can try them in real sessions before the clinic standardizes. The pens and needles are supplied direct from the manufacturer in Türkiye, with no importer in between, and distributors who want to carry the line can apply for a dealer account.